A low-cost disposal pencil electrode for cyclic
voltammetry analysis
Pooja Chandravanshi1, Ankita Tejwani1,
Tushar Kant2, Manas Kanti Deb1, and Kamlesh Shrivas1*
1School
of Studies in Chemistry, Pt. Ravishankar Shukla University, Raipur-492010, CG,
India
2Shaheed
Kawasi Rodda Pedda, Govt. College Kuakonda, Dantewada-494552, CG, India
Abstract.
This study explores the use
of disposable pencil electrodes in cyclic voltammetry (CV) and compares their
performance to traditional glassy carbon electrodes. We conducted CV tests to
evaluate different pencil grades, HB, 2B, 4B, 6B, 8B, and 10B along with glassy
carbon to assess response time, peak current, and electrode stability during
the analysis of potassium ferrocyanide. The results showed that the 4B pencil
outperformed other grades and the glassy carbon electrode, with faster response
times, higher peak currents, and greater stability. Notably, the 4B pencil's
anodic and cathodic currents were higher at 25 mV/s scan rate. This work
demonstrates that 4B pencils can serve as cost-effective and reliable
alternatives to glassy carbon electrodes in electrochemical applications,
offering an accessible option for potassium ferrocyanide analysis and broader
electrochemical studies.
Keywords: Pencil electrode, electrochemical response,
potassium ferrocyanide, cyclic voltammetry, electrode material.
Introduction
Electrochemical
reactions involve the transfer of electrons between chemical species, leading
to changes in their oxidation states. These reactions underpin key processes
like energy storage, corrosion, and catalysis. They occur within
electrochemical cells, where electrodes drive redox reactions, converting
chemical energy into electrical energy, or vice versa. Voltammetry is a
technique used to study how current response to applied potential in these
systems. It encompasses a variety of methods, including cyclic voltammetry
(CV), differential pulse voltammetry (DPV), square wave voltammetry (SWV), and
linear sweep voltammetry (LSV). These techniques provide valuable data on the
kinetics of electron transfer among electroactive species in solution (Elgrishi, Rountree et al. 2018, Rana, Baig et al. 2019).
In CV, the
potential applied to an electrode is linearly cyclically swept between two
fixed values while measuring the current flow. The resulting plot of current
versus potential is known as a cyclic voltammogram. CV is a versatile technique
used across various fields like analytical chemistry, materials science,
electroplating, and the development of electrochemical sensors and biosensors (Tavares and Barbeira 2008, King, Friend et al. 2010,
Marken, Neudeck et al. 2010, Dilgin, Kızılkaya et al. 2012). It provides
crucial insights into the behavior of electroactive species in solution. A
typical CV setup includes a potentiostat to control the potential applied to
the working electrode and a three-electrode cell comprising a working
electrode, a reference electrode, and a counter electrode. The working
electrode is where the electrochemical reaction occurs; it is typically made of
conductive materials like platinum, gold, or glassy carbon electrode (GCE) and
can be customized with various materials such as nanoparticles, polymers, or
enzymes for specific purposes (Mabbott 1983,
Rusling and Suib 1994, Menshykau, Streeter et al. 2008, Chan, Koch et al. 2017,
Cook, Ko et al. 2023). A
GCE is a type of carbon-based electrode with a highly ordered structure that
resembles glass in terms of hardness and resistance. It offers high chemical
stability, low porosity, and good conductivity, making it ideal for
electrochemical applications. However, GCE are costly, rigid, and have limited
surface area compared to other carbon-based electrodes, such as graphite pencil
electrodes. The reference electrode, often a saturated calomel electrode
(SCE) or a silver/silver chloride electrode (Ag/AgCl), maintains a stable
potential against which the working electrode's potential is compared. The
counter electrode provides a current pathway, completing the electrical circuit
(Evans, O'Connell et al. 1983, Kissinger and
Heineman 1983, Grenthe, Stumm et al. 1992, Hagfeldt and Graetzel 1995, 2013,
Lakhera, Chaudhary et al. 2022). This setup allows researchers to study
redox processes, evaluate material properties, and develop applications for
sensors and biosensors.
Graphite, known for its unique combination of metallic and
nonmetallic properties, serves as an excellent electrode material. Graphite is a naturally
occurring carbon mineral with a hexagonal lattice structure, commonly used in
various industries due to its unique properties. Graphite is non-toxic,
chemically stable, and resistant to corrosion, oxidation, and most chemical
reactions. Graphite is an excellent conductor of electricity and heat, used in
electronics for electrodes, brushes, and collectors. Its high thermal
conductivity and anisotropic structure, which varies in strength depending on
the measurement direction, add to its versatility. Graphite pencil electrode (GPE) are popular for their low cost, ease of
use, commercial availability, and disability (Annu,
Sharma et al. 2020).
GPE have been
utilized in a range of innovative applications. For instance, Purushottam et
al. devised a method for analyzing chlorpromazine using GPE. Kevin C et al.
used GPE for the detection of Pb2+ in anodic stripping voltammetry,
demonstrating an eco-friendly, cost-effective alternative for electrochemical
devices. Tadi et al. employed molecularly imprinted polymer (MIP) technology
for sulfanilamide (SN) sensing, using in-situ electro-polymerization of pyrrole
on pencil graphite electrodes. Heydari et al. developed a hydrazine sensor by
depositing copper nanostructures (CuNs) on modified GPE. These examples
illustrate the versatility and cost-effectiveness of GPE in a wide range of
electrochemical applications (Annu, Sharma et al. 2020, da Silva, Paschoalino et al. 2020).
This study aims to assess the suitability of disposable
pencil electrodes for CV compared to GCE. Evaluating different pencil grades
alongside glassy carbon, it identifies 4B pencils as superior, offering faster
response times, higher peak currents, and greater stability, suggesting their
potential as cost-effective alternatives in electrochemical analysis.
Materials and Methods
Chemical and reagent
All the chemicals used were of analytical reagent (AR)
grade. Potassium ferrocyanide K4[Fe (CN)6], (98.0%),
potassium chloride (KCl, 99.5%), were purchased from HIMEDIA (Mumbai, India).
Double distilled water (DW) was used for the preparation of standard solutions
and the cyclic voltammetry (CV) analysis. Phosphate buffer solutions (PBS) of
pH 7.0 and 9.0 were prepared from NaH2PO4 (99.0%) and Na2HPO4
(98.0%) obtained from HIMEDIA (Mumbai, India).
Apparatus
A digital multimeter was used for the measurement of
resistance in different types of HB pencils and Potentiostat (Squidstat Solo,
Admiral Instrument, USA) was employed for CV measurements of K4[Fe (CN)6].
Preparation of different pencil electrodes
To prepare pencil
electrodes for electrochemical applications, the pencils with different
graphite concentrations, such as HB, 2B, 4B, 6B, 8B, and 10B were selected for
present investigations (Nixon
and Parry 1968, Kopelevich and Esquinazi 2007). These grades vary in hardness and
brightness, with H indicating hardness and B indicating softness.
The sharp blade was used to carefully remove of wooden casing to expose the
graphite lead. To control the exposed area, tightly wrap the remaining pencil
with Teflon tape, leaving only the intended electrode surface open to the
solution. After preparation, test the pencil electrode in a standard solution
to confirm its stability and functionality, adjusting as needed to ensure a
consistent and reliable electrode for your electrochemical applications.
Electrochemical analysis of potassium ferrocyanide
using GPE
For the electrochemical measurement
of K4[Fe (CN)6] using CV with pencil electrodes, follow
this methodology. The select pencils with varying graphite grades, including
HB, 2B, 4B, 6B, 8B, and 10B, to use as working electrodes. Commercially
obtained electrodes, specifically a silver/silver chloride (Ag/AgCl) electrode
as the reference and a platinum electrode as the counter, should also be
employed. In the electrochemical
cell, prepare a phosphate-buffered saline (0.2M PBS, pH=7.0)
solution with potassium chloride (KCl, 1.0 M) as the
supporting electrolyte. Insert the pencil electrode into the electrolyte
solution alongside the reference and counter electrodes. Ensure that only the
desired length of the pencil electrode's graphite is exposed to the solution to
control the electrode's surface area. The set up the potentiostat for CV,
ensuring correct connections to the working, reference, and counter electrodes.
The CV measurements were done by sweeping the potential linearly within a
predetermined range and at a specific scan rate to observe the redox behavior
of K4[Fe (CN)6]. Record the resulting CV to analyze the
electrochemical responses and determine the electrochemical characteristics of
the pencil electrodes.
Result and Discussion
Electrochemical study of K4[Fe (CN)6] in
CV using GPE
In this study, we
explored the electrochemical activity of different pencil electrodes in the CV
measurement of K4[Fe (CN)6]. Various types of graphite
pencils, HB, 2B, 4B, 6B, 8B, and 10B were used as working electrodes, as
depicted in Fig. 1 (a and b).
Additionally, commercially sourced Ag/AgCl and platinum electrodes were used as
the reference and counter electrodes, respectively. To conduct the experiment, a three-electrode
setup was employed, consisting of a graphite pencil electrode as the working
electrode, an Ag/AgCl electrode as the reference, and a platinum electrode as
the counter. This setup was connected to a potentiostat from Admiral
Instruments, allowing precise control and measurement of the applied potential
and resultant current, shown in Fig. 2.
The electrochemical cell contained a 0.1 M solution of potassium ferrocyanide K4[Fe
(CN)6], with 1 M KCl as the supporting electrolyte and 0.2 M PBS to
maintain a stable pH environment. The working electrode was immersed in the
solution along with the reference and counter electrodes, ensuring proper
contact and positioning. CV measurements were performed by sweeping the
potential across a specified range while recording the resulting current.
Fig. 1. (a and b) illustrate the different types of
graphite pencils electrodes (HB, 2B, 4B,
6B, 8B, and 10B)
Fig. 2.
Scheme for three electrode systems for CV measurement of (K4[Fe(CN)6]).
Fig. 3 shows the anodic
and cathodic current peaks in CV measurements of K4[Fe
(CN)6]. This
technique was used to evaluate the redox behavior of the ferrocyanide compound
and to assess the performance of different pencil electrodes. In CV, the redox reactions occurring at the
cathode and oxidation at anode. For the given reaction involving (𝐹𝑒(𝐶𝑁)6)4−
and (Fe (CN)6)3−, the reduction at the cathode and
oxidation at the anode, respectively can be described by the following
half-reactions:
Reduction at cathode:
When the potential
was changed from 1.2 V to −0.75 V, the reduction occurred at the cathode at 0.1
V. In CV, the reduction reaction
at the cathode typically occurs at potentials slightly below the standard
reduction potential due to the applied potential and reaction kinetics.
(𝐹𝑒(𝐶𝑁)6)4−+𝑒−→(𝐹𝑒(𝐶𝑁)6)3−
Oxidation at anode:
When the potential
was changed from −0.75 V to 1.2 V, the oxidation occurred at the anode at 0.6 V.
In CV, the oxidation reaction at
the anode typically occurs at potentials slightly above the standard reduction
potential due to the applied potential and reaction kinetics.
(𝐹𝑒(𝐶𝑁)6)3−→(𝐹𝑒(𝐶𝑁)6)4−+𝑒−
In summary, in CV, reduction at the cathode typically occurs
at potentials slightly below the standard reduction potential, and oxidation at
the anode typically occurs at potentials slightly above the standard reduction
potential, due to the applied potential and reaction kinetics.
Potential (Volt)
Fig. 3. CV
peaks of using of (K4[Fe (CN)6] using 4B pencil electrode
as a working electrode at 25 mV/s
Electrochemical
performance of GPE for measurement of K4[Fe (CN)6] in CV
Anodic and
cathodic currents in CV refer to the flow of electrons during oxidation and
reduction reactions, respectively. Anodic current is generated when the
potential applied to the working electrode causes oxidation, driving electrons
from the electrode to the solution. Cathodic current is observed when the
potential drives electrons from the solution to the electrode, indicating
reduction (Tasić, Mihajlović et al. 2022).
Based on the given Table 1, the
trends in anodic and cathodic currents for different types of pencil electrodes
at various scan rates (25, 50, 100 mV/s) can be analyzed.
The anodic current
represents the flow of electrons from the electrode to the solution during
oxidation. In general, the anodic current increases as the scan rate increases,
indicating that higher scan rates lead to more rapid oxidation (Deepa, Madhu et al. 2021). Among the different
pencil types, 4B shows the highest anodic current at the 25 mV/s scan rate
(0.0051 A), with a similar trend at 50 mV/s (0.007 A). However, at the 100 mV/s
scan rate, 4B's anodic current drops significantly (0.0013 A). This fluctuation
might suggest a change in conductivity or resistance at higher scan rates.
Thus, 25 mV/s scan rate was found for analysis of K4[Fe (CN)6]
using 4B pencil as a working electrode in anodic current analyses.
Further, cathodic
reflects the flow of electrons from the solution to the electrode during
reduction. Similar to the anodic current, cathodic currents generally increase
with higher scan rates (Shama, Asir et al. 2020).
At 25 mV/s, the cathodic current for 4B (-0.0054 A) is much higher than other
pencils, indicating a strong reduction process. However, at 100 mV/s, the
cathodic current for 4B drops significantly (-0.0101 A), suggesting potential
changes in electrode behavior at higher scan rates. Herein, also 25 mV/s scan
rate was found for analysis of K4[Fe (CN)6] using 4B pencil as a
working electrode in cathodic analyses. Fig. 4. Shows the anodic and cathodic currents at at different
scanning rate (a)25mV/s, (b) 50mV/s, and (c) 100mV/s using different types of GPE in CV analsyes.
Overall, the table
illustrates that scan rates and electrode type play crucial roles in
determining the anodic and cathodic currents. It also suggests that 4B is the
most effective pencil type at 25 mV/s scan rate.
Table 1. Anodic current and cathodic current is
obtained in CV at scanning rates of 25, 50, 100 mV/s.
|
Type of Pencil
|
Scanning rate, 25 mV/s
|
Scanning rate, 50 mV/s
|
Scanning rate, 100 mV/s
|
Anodic current
|
Cathodic
current
|
Anodic current
|
Cathodic
current
|
Anodic current
|
Cathodic
current
|
HB
|
0.0009
|
-0.0003
|
0.0008
|
-0.0002
|
0.0028
|
-0.0008
|
2B
|
0.0013
|
-0.0006
|
0.0012
|
-0.0005
|
0.0015
|
-0.0006
|
4B
|
0.0051
|
-0.0054
|
0.007
|
-0.0063
|
0.0013
|
-0.0101
|
6B
|
0.0023
|
-0.0009
|
0.0025
|
-0.0015
|
0.0045
|
-0.0028
|
8B
|
0.0021
|
-0.0006
|
0.0021
|
-0.0009
|
0.0042
|
-0.0017
|
10B
|
0.002
|
-0.0006
|
0.0028
|
-0.0014
|
0.0032
|
-0.0016
|
Fig. 4. Anodic and cathodic currents at different scanning rates (a)
25mV/s, (b) 50 mV/s,
and (c) 100 mV/s using different types of
GPE in CV analysis
Electrochemical
performance of GPE with size, diameter, and resistance in the measurement of K4[Fe
(CN)6] in CV
Anodic and
cathodic currents in electrochemical studies represent the flow of electrons
during oxidation and reduction processes, respectively. In CV, the anodic
current refers to electron flow from the electrode to the solution (oxidation),
while the cathodic current is electron flow from the solution to the electrode
(reduction) (Devkota, Chuangchote et al. 2021,
Kayali, Shama et al. 2023). The values from your given table reflect
these currents at a scanning rate of 25 mV/s for various pencil types, Shown in
Table 2.
This current
occurs when the electrode facilitates oxidation reactions. The higher the
anodic current, the greater the oxidation activity at the electrode. According
to table, the 4B pencil exhibits the highest anodic current (0.0051 A),
indicating it is the most efficient for oxidation reactions at this scan rate.
The other pencils show lower anodic currents, with HB being the lowest at
0.0009 A. This trend suggests that 4B offers better conductive properties or a
more favorable structure for electron flow during oxidation.
This current is
observed during reduction reactions, when electrons move from the solution to
the electrode. As with anodic currents, the 4B pencil leads with the highest
cathodic current (-0.00542 A), indicating strong reduction activity. The other
pencils display lower cathodic currents, with HB again being the lowest
(-0.00026 A), suggesting that 4B's structure and composition are optimal for
reduction reactions.
The table also
provides resistance, radius, and diameter information, which can influence the
anodic and cathodic currents. Notably, the resistance tends to decrease with
increasing diameter and radius, with the exception of 2B, which has a higher
resistance. This decrease in resistance may be why 4B shows higher current
values the lower the resistance, the easier it is for electrons to flow,
resulting in stronger anodic and cathodic currents.
In summary, the
table indicates that among the pencils tested, the 4B type has the highest
anodic and cathodic currents at a 25 mV/s scan rate, suggesting it is the most
efficient for both oxidation and reduction reactions. This is likely due to a
combination of its optimal radius, diameter, and lower resistance compared to
other pencil types.
Table 2. Type of pencil, radius, diameter, resistance, anodic
current and cathodic current at 25mV/s scanning rate.
|
Type of
pencil
|
Radius
(cm)
|
Diameter (cm)
|
Resistance (Ω)
|
Anodic Current (A)
|
Cathodic current (A)
|
HB
|
0.11
|
0.22
|
0.019
|
0.0009
|
-0.00026
|
2B
|
0.14
|
0.28
|
0.022
|
0.0013
|
-0.00066
|
4B
|
0.18
|
0.36
|
0.015
|
0.0051
|
-0.00542
|
6B
|
0.19
|
0.38
|
0.011
|
0.0023
|
-0.00094
|
8B
|
0.20
|
0.40
|
0.008
|
0.0021
|
-0.00065
|
10B
|
0.21
|
0.42
|
0.006
|
0.002
|
-0.00064
|
Comparision of performance of GCE and
GPE in CV analyses
GPE
can produce higher anodic and cathodic currents for potassium ferrocyanide (K4[Fe
(CN)6]) in CV compared to GCEs due to their unique structure and
composition, shown in Fig. 5.
Graphite, with its layered structure, provides a high density of delocalized
electrons, enhancing electron transfer efficiency during redox reactions. This
property can lead to greater interaction with the electroactive species in CV.
Additionally, the larger surface area and flexibility of GPEs, compared to the
rigid structure of GCEs, offer more contact points with the electrolyte,
facilitating greater electron flow. The combination of these factors can result
in higher anodic and cathodic currents with GPE compared to GCEs in CV analysis
of potassium ferrocyanide.
Fig. 5. CV peaks of K4[Fe (CN)6] in
CV when (a) Glassy carbon electrode (b) 4B Pencil electrode
Conclusions
This research focused on the use of GPE in CV, comparing
their performance to traditional GCEs in
the analysis of potassium ferrocyanide (K4[Fe(CN)6]). The
study revealed that GPE, particularly the 4B grade, exhibited superior
electrochemical characteristics, with higher anodic and cathodic currents,
faster response times, and greater stability compared to GCEs. The novelty lies
in demonstrating that GPE are a cost-effective and flexible alternative to GCEs
for electrochemical applications. These findings suggest that GPE can be
valuable in diverse electrochemical studies, offering an accessible,
disposable, and efficient option for researchers and practitioners.
Acknowledges
Kamlesh Shrivas
acknowledged the Department of Science &
Technology (DST)-Promotion of University Research and
Scientific Excellence (SR/PURSE/2022/145) and Chhattisgarh Council of
Science and Technology (Ref. No: CCOST/EMR/2023), Raipur for providing the
financial assistance.
References
Annu,
S. Sharma, R. Jain and A. N. Raja (2020).
Review—Pencil Graphite Electrode: An Emerging Sensing Material. Journal of The Electrochemical Society 167(3): 037501.
Chan,
G. G., C. M. Koch and L. H. Connors (2017).
Blood Proteomic Profiling in Inherited (ATTRm) and Acquired (ATTRwt)
Forms of Transthyretin-Associated Cardiac Amyloidosis. Journal of Proteome Research 16(4): 1659-1668.
Cook,
J. B., J. S. Ko, T. C. Lin, D. D. Robertson, H.-S. Kim, Y. Yan, Y. Yao, B. S.
Dunn and S. H. Tolbert (2023). Ultrafast
Sodium Intercalation Pseudocapacitance in MoS2 Facilitated by Phase Transition
Suppression. ACS Applied Energy
Materials 6(1): 99-108.
da
Silva, A. D., W. J. Paschoalino, J. P. V. Damasceno and L. T. Kubota
(2020). Structure, Properties, and
Electrochemical Sensing Applications of Graphene-Based Materials. ChemElectroChem 7(22): 4508-4525.
Deepa,
H. A., G. M. Madhu and B. E. K. Swamy (2021).
Evaluation of performance characteristics of nano TiO2 and TiO2-ZnO
composite for DSSC applications and electrochemical determination of potassium
ferrocyanide using cyclic voltammetry.
Materials Research Express 8(12):
125004.
Devkota,
M., S. Chuangchote, C. La-o-vorakiat, P. Lertsathitphong, B. Lertanantawong, M.
Somasundrum and W. Surareungchai (2021).
Photoelectrochemical reduction rate of ferricyanide at different TiO2
forms: comparison of SECM and cyclic voltammetric results. Journal of Solid State Electrochemistry 25(5): 1691-1698.
Dilgin,
Y., B. Kızılkaya, B. Ertek, F. Işık and D. G. Dilgin (2012). Electrocatalytic oxidation of sulphide using
a pencil graphite electrode modified with hematoxylin. Sensors and Actuators B: Chemical 171-172: 223-229.
Elgrishi,
N., K. J. Rountree, B. D. McCarthy, E. S. Rountree, T. T. Eisenhart and J. L.
Dempsey (2018). A Practical Beginner’s
Guide to Cyclic Voltammetry. Journal of
Chemical Education 95(2): 197-206.
Evans,
D. H., K. M. O'Connell, R. A. Petersen and M. J. Kelly (1983). Cyclic voltammetry. Journal of Chemical Education 60(4): 290.
Grenthe,
I., W. Stumm, M. Laaksuharju, A. C. Nilsson and P. Wikberg (1992). Redox potentials and redox reactions in deep
groundwater systems. Chemical Geology 98(1): 131-150.
Hagfeldt,
A. and M. Graetzel (1995). Light-Induced
Redox Reactions in Nanocrystalline Systems.
Chemical Reviews 95(1):
49-68.
Kayali,
D., N. A. Shama, S. Asir and K. Dimililer (2023). Machine learning-based models for the
qualitative classification of potassium ferrocyanide using electrochemical
methods. The Journal of Supercomputing 79(11): 12472-12491.
King,
D., J. Friend and J. Kariuki (2010).
Measuring Vitamin C Content of Commercial Orange Juice Using a Pencil
Lead Electrode. Journal of Chemical
Education 87(5): 507-509.
Kissinger,
P. T. and W. R. Heineman (1983). Cyclic
voltammetry. Journal of Chemical
Education 60(9): 702.
Kopelevich,
Y. and P. Esquinazi (2007). Graphene
Physics in Graphite. Advanced Materials 19(24): 4559-4563.
Lakhera,
P., V. Chaudhary, A. Jha, R. Singh, P. Kush and P. Kumar (2022). Recent developments and fabrication of the
different electrochemical biosensors based on modified screen printed and
glassy carbon electrodes for the early diagnosis of diverse breast cancer
biomarkers. Materials Today Chemistry 26: 101129.
Mabbott,
G. A. (1983). An introduction to cyclic
voltammetry. Journal of Chemical
Education 60(9): 697.
Marken,
F., A. Neudeck and A. M. Bond (2010). Cyclic Voltammetry. Electroanalytical
Methods: Guide to Experiments and Applications. F. Scholz, A. M. Bond, R. G.
Compton et al. Berlin, Heidelberg, Springer Berlin Heidelberg: 57-106.
Menshykau,
D., I. Streeter and R. G. Compton (2008).
Influence of Electrode Roughness on Cyclic Voltammetry. The Journal of Physical Chemistry C 112(37): 14428-14438.
Nixon,
D. E. and G. S. Parry (1968). Formation
and structure of the potassium graphites.
Journal of Physics D: Applied Physics 1(3): 303.
Rana,
A., N. Baig and T. A. Saleh (2019).
Electrochemically pretreated carbon electrodes and their
electroanalytical applications – A review.
Journal of Electroanalytical Chemistry 833: 313-332.
Rusling,
J. F. and S. L. Suib (1994).
Characterizing Materials with Cyclic Voltammetry. Advanced Materials 6(12): 922-930.
Shama,
N. A., S. Asir, K. Dimililer and M. Özsöz (2020). Potassium Ferro cyanide
electrochemically detected by Differential Pulse and Square Wave Voltammetry in
a Competition using Gradient Boosting as Machine Learning Algorithm. 2020 4th
International Symposium on Multidisciplinary Studies and Innovative
Technologies (ISMSIT).
Tasić,
Ž. Z., M. B. P. Mihajlović, M. B. Radovanović, A. T. Simonović, D. V. Medić and
M. M. Antonijević (2022).
Electrochemical determination of L-tryptophan in food samples on
graphite electrode prepared from waste batteries. Scientific Reports 12(1): 5469.
Tavares,
P. H. C. P. and P. J. S. Barbeira (2008).
Influence of pencil lead hardness on voltammetric response of graphite
reinforcement carbon electrodes. Journal
of Applied Electrochemistry 38(6):
827-832.